Option
Description
Against Arrow The majority of flow through the sensor is opposite to the Flow arrow on the sensor. Actual
forward flow is processed as reverse flow.
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow direction or the
setting of
Sensor Flow Direction Arrow
.
Sensor Flow Direction Arrow
controls only whether actual flow is
processed as forward flow or reverse flow.
Related information
Configure mA Output Direction
Configure Frequency Output Direction
Configure Discrete Output Source
Configure totalizers and inventories
Effect of Sensor Flow Direction Arrow on digital communications
4.2 Configure mass flow measurement
The mass flow measurement parameters control how mass flow is measured and reported. The mass total
and mass inventory are derived from the mass flow data.
4.2.1 Configure Mass Flow Measurement Unit
Display
Menu
>
Configuration
>
Process Measurement
>
Flow Variables
>
Mass Flow Settings
>
Units
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
>
Mass Flow Rate Unit
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Flow
>
Mass Flow Unit
Mass Flow Measurement Unit
specifies the unit of measure that will be used for the mass flow rate. The
default unit used for mass total and mass inventory is derived from this unit.
Procedure
Set
Mass Flow Measurement Unit
to the unit you want to use.
•
Default: g/sec (grams per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for
Mass Flow Measurement Unit
, plus one
user-defined special measurement unit. Different communications tools may use different labels for the
units.
Unit description
Label
Display
ProLink III
Field Communicator
Grams per second
gram/s
g/sec
g/s
Configure process measurement
Configuration and Use Manual
March 2019
MMI-20025166
32
Micro Motion Model 5700 Transmitters with Configurable Outputs