Options for Discrete Output Source
Option
Label
State DO voltage
Display
PLIII
FC
Web browser
Enhanced
Event 1–5
Basic Event x
Enhanced
Event x
Discrete
Event x
Enhanced
Event x
ON
•
Internally powered:
24 VDC
•
Externally powered: Site-
specific
OFF
0 V
Enhanced
Event x
Event x
Event x
N/A
ON
•
Internally powered:
24 VDC
•
Externally powered: Site-
specific
OFF
0 V
Flow Rate
Switch
Flow Rate
Switch
Flow Switch
Indicator
Flow Switch
Flow Rate
Switch
ON
•
Internally powered:
24 VDC
•
Externally powered: Site-
specific
OFF
0 V
Forward/
Reverse
Indicator
Flow
Direction
Forward
Reverse
Indicator
Forward/
Reverse
Flow Direction
Forwa
rd
flow
0 V
Rever
se
flow
•
Internally powered:
24 VDC
•
Externally powered: Site-
specific
Calibration in
Progress
Zero in
Progress
Calibration in
Progress
Cal in
Progress
Zero in
Progress
ON
•
Internally powered:
24 VDC
•
Externally powered: Site-
specific
OFF
0 V
Fault
Fault
Fault
Indication
Fault
Condition
Present
Fault
ON
•
Internally powered:
24 VDC
•
Externally powered: Site-
specific
OFF
0 V
Meter
Verification
Failure
Meter
Verification
Fail
Meter
Verification
Failure
SMV Fail
Meter
Verification Fail
ON
•
Internally powered:
24 VDC
•
Externally powered: Site-
specific
OFF
0 V
Configuration and Use Manual
Integrate the meter with the control system
MMI-20025166
March 2019
Configuration and Use Manual
159