B.2.2 Make a service port connection from ProLink III to the
transmitter
CAUTION
If the transmitter is in a hazardous area, do not open the wiring compartment while the transmitter is
powered up. Opening the wiring compartment while the transmitter is powered up could cause an
explosion. To connect to the transmitter in a hazardous environment, use a connection method that does
not require opening the wiring compartment.
Prerequisites
•
Ensure the transmitter service port is enabled.
•
Obtain a USB type A to type A cable.
Important
The USB cable should be no greater than 1 meter in length.
Procedure
1. Insert one end of the USB cable into the USB port on your PC.
2. Open the wiring compartment on the transmitter, and insert the other end of the USB cable into the
service port on the transmitter.
Configuration and Use Manual
Using ProLink III with the transmitter
MMI-20025166
March 2019
Configuration and Use Manual
317