Recommended actions
Check for air entrainment, tube fouling, flashing, or tube damage.
17.13.44 Batcher Not Configured
Alert
Configuration Error
Cause
One or more of the following has occurred:
•
The batcher application is disabled
•
No flow source has been configured for the batcher application
•
The batch target is 0
•
No discrete output is assigned to batch control.
Recommended actions
Verify the configuration of the batcher application and related parameters.
17.13.45 Batch Time Out
Alert
Process Aberration
Cause
The batch exceeded
Maximum Batch Time
, and ended before the target was reached.
Recommended actions
Verify the configuration of the batcher application and related parameters.
17.13.46 Clock Failure
Alert
Data Loss Possible
Cause
The transmitter's real-time clock is not incrementing.
Recommended actions
Contact customer service.
17.13.47 Core Processor Update Failed
Alert
Configuration Error
Configuration and Use Manual