Figure 3-2: Removing the transmitter housing cover
3. Using a fine-pointed tool, move the switch to the desired position.
4. Replace the transmitter housing cover.
5. If necessary, power up the transmitter.
3.1.3 Enable or disable the service port
Display
Menu
>
Configuration
>
Security
>
Service Port
ProLink III
Not available
Field Communicator
Configure
>
Manual Setup
>
Security
>
Enable/Disable Service Port
The service port is enabled by default, so you can use it for transferring files or connect to it with ProLink III. If
you want to completely prevent it from being used, you can disable it.
Note
Enabling or disabling the service port will not take effect until power has been cycled to the transmitter.
CAUTION
Do not use the service port if the transmitter is in a hazardous area. To use the service port, you must open
the transmitter wiring compartment. Opening the wiring compartment in a hazardous area, while the
transmitter is powered up, can cause an explosion.
Configuration and Use Manual
Introduction to configuration and commissioning