Method
Description
Setup
Polling
The meter polls an external
device for temperature data.
This data will be available in
addition to the internal
temperature data.
a. Choose
Menu
>
Configuration
>
Process Measurement
>
Temperature
.
b. Set
External Temperature
to On.
c. Choose
Poll External Device
.
d. Select
Polled Variable 1
or
Polled Variable 2
.
e. Set
Variable
to External Temperature.
f. Set
Polling Control
to Poll as Primary or Poll as Secondary.
Option
Description
Poll as Primary
No other HART masters will be on the
network. The Field Communicator is
not a HART master.
Poll as Secondary Other HART masters will be on the
network. The Field Communicator is
not a HART master.
g. Set
External Device Tag
to the HART tag of the external
temperature device.
mA Input
An external device sends
temperature data to the meter
via the mA Input. This data will
be available in addition to the
internal temperature data.
a. Choose
Menu
>
Configuration
>
Process Measurement
>
Temperature
.
b. Set
External Temperature
to On.
c. Choose
Menu
>
Configuration
>
Inputs/Outputs
>
Channel
D
.
d. Set
I/O Type
to MAI.
e. Set
Power
appropriately.
f. Choose
I/O Settings
.
g. Set
Assignment
to External Temperature.
h. Set
Lower Range Value
and
Upper Range Value
appropriately.
Digital
communications
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal temperature data.
a. Choose
Menu
>
Configuration
>
Process Measurement
>
Temperature
.
b. Set
External Temperature
to On.
c. Perform the necessary host programming and
communications setup to write temperature data to the
transmitter at appropriate intervals.
Postrequisites
Choose
Menu
>
Service Tools
>
Service Data
>
View Process Variables
and verify the value for External
Temperature.
Configuration and Use Manual
Configure process measurement applications
MMI-20025166
March 2019
Configuration and Use Manual
77