Replication files
Contain all parameters except the device-specific parameters, e.g., calibration factors or
meter factors. They are used to replicate the transmitter configuration to other devices.
The
.xfer
extension is used to identify replication files.
Tip
You can use a saved configuration file to change the nature of the transmitter quickly. This might be
convenient if the transmitter is used for different applications or different process fluids.
Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually
or by groups. Therefore, you can use this format for both backup and replication.
Procedure
•
To save the current configuration to the transmitter's SD card:
a) Choose
Device Tools
>
Configuration Transfer
>
Save Configuration
.
b) Select On my 5700 Device Internal Memory and click
Next
.
c) Click
Save
.
d) Enter the name for this configuration file.
e) Set the file type.
—
To save a backup file, set the file type to Backup.
—
To save a replication file, set the file type to Transfer.
f) Click
Save
.
The configuration file is saved to the transmitter's SD card as
yourname
.spare
or
yourname
.xfer
.
•
To save the current configuration to your PC, in Model 5700 format:
a) Choose
Device Tools
>
Configuration Transfer
>
Save Configuration
.
b) Select On my computer in 5700 device file format and click
Next
.
c) Click
Save
.
d) Browse to the desired location, then enter the name for this configuration file.
e) Set the file type.
—
To save a backup file, set the file type to Backup.
—
To save a replication file, set the file type to Transfer.
f) Click
Save
.
The configuration file is saved to the specified location as
yourname
.spare
or
yourname
.xfer
.
•
To save the current configuration to your PC, in ProLink III format:
a) Choose
Device Tools
>
Configuration Transfer
>
Save Configuration
.
b) Select On my computer in ProLink III file format and click
Next
.
c) Click
Save
.
Introduction to configuration and commissioning