5.2.2 Set up concentration measurement using the display
This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Restriction
This section does not cover building a concentration matrix. See the
Micro Motion Enhanced Density Application
Manual
for detailed information on building a matrix.
Enable the concentration measurement application using the display
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Procedure
1. Choose
Menu
>
Configuration
>
Process Measurement
.
2. Choose
Flow Variables
>
Volume Flow Settings
and ensure that
Flow Type
is set to Liquid.
3. Return to the
Process Measurement
menu.
4. If the API referral application is displayed in the menu, choose
API Referral
and ensure that
Enabled/
Disabled
is set to Disabled.
The concentration measurement application and the API referral application cannot be enabled
simultaneously.
5. If the advanced phase measurement application is displayed in the menu, choose
Advanced Phase
Measurement
>
Application Setup
and ensure that
Enabled/Disabled
is set to Disabled.
The concentration measurement application and the advanced phase measurement application
cannot be enabled simultaneously.
6. Enable concentration measurement.
a) Choose
Menu
>
Configuration
>
Process Measurement
>
Concentration Measurement
.
b) Set
Enabled/Disabled
to Enabled.
Load a concentration matrix from a USB drive using the display
At least one concentration matrix must be loaded into one of the six slots on your transmitter. You can load
up to six matrices into slots. You can also copy matrices to the transmitter's SD card, and load them into slots
at a later time.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not
need to load any matrices.