6 Configure advanced options for process
measurement
6.1 Configure Response Time
Display
Menu
>
Configuration
>
Process Measurement
>
Response Time
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Response Time
Field Communicator
Not available
Response Time
controls the speed of various internal processes that are involved in retrieving electronic data
from the sensor and converting it to process data.
Response Time
affects all process and diagnostic variables.
Restriction
Response Time
is configurable only if you are using the enhanced core processor. If you are using the
standard core processor,
Response Time
is set to Low Filtering and cannot be changed.
Procedure
Set
Response Time
as desired.
Option
Description
Normal
Appropriate for typical applications.
High Filtering Slower response. Appropriate for applications with significant amount of entrained gas or
process noise.
Low Filtering Fastest response. Appropriate for proving or filling applications.
Service
Do not select unless directed by Micro Motion personnel.
6.2 Detect and report two-phase flow
Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of process control issues.
The transmitter provides two methods to detect and report or respond to two-phase flow.
6.2.1 Detect two-phase flow using density
Display
Menu
>
Configuration
>
Process Measurement
>
Density
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Density
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
Density
>
Slug Low Limit
Configure
>
Manual Setup
>
Measurements
>
Density
>
Slug High Limit
Configure
>
Manual Setup
>
Measurements
>
Density
>
Slug Duration
Configuration and Use Manual
Configure advanced options for process measurement
MMI-20025166
March 2019
Configuration and Use Manual
105