Procedure
1. Choose
Menu
>
Configuration
>
Process Measurement
>
Concentration Measurement
>
Configure
Matrix
.
2. Select the matrix that you want to modify.
3. Set
Extrapolation Limit
to the point, in percent, at which an extrapolation alert will be posted.
4. Choose
Menu
>
Configuration
>
Process Measurement
>
Concentration Measurement
>
Configure
Application
>
Extrapolation Alerts
.
5. Enable or disable the high and low limit alerts for temperature and density as desired.
Extrapolation alerts in action
If
Extrapolation Limit
is set to 5%,
High Limit (Temp)
is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using the display
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose
Menu
>
Configuration
>
Process Measurement
>
Concentration Measurement
>
Configure
Application
.
2. Set
Active Matrix
to the matrix you want to use.
5.2.3 Set up concentration measurement using ProLink III
This section guides you through the tasks required to set up, configure, and implement concentration
measurement.
Enable the concentration measurement application using ProLink III
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
•
The concentration measurement application must be licensed on your transmitter.
•
The concentration measurement application cannot be enabled at the same time as the API referral
application or the Advanced Phase Measurement application. They must be disabled first.
Procedure
1. Choose
Device Tools
>
Configuration
>
Process Measurement
>
Flow
and ensure that
Volume Flow
Type
is set to Liquid Volume.
2. Choose
Device Tools
>
Configuration
>
Transmitter Options
.
3. Set
Concentration Measurement
to Enabled and click
Apply
.
Configuration and Use Manual
Configure process measurement applications
MMI-20025166
March 2019
Configuration and Use Manual
79