6
Integrate the meter with the control
system
Topics covered in this chapter:
•
Configure the transmitter channels
•
•
Configure the frequency output
•
•
•
Configure digital communications
6.1
Configure the transmitter channels
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
IO
>
CH B
ProLink II
ProLink
>
Configuration
>
Frequency/Discrete Output
ProLink III
Device Tools
>
Configuration
>
I/O
>
Channels
Field Communicator
Configure
>
Manual Setup
>
Inputs/Outputs
>
Channels
>
Channel B
Overview
You can configure Channel B on your transmitter to operate as a frequency output or a
discrete output. The channel configuration must match the wiring at the transmitter
terminals.
Prerequisites
To avoid causing process errors:
•
Configure the channels before configuring the outputs.
•
Before changing the channel configuration, ensure that all control loops affected by
the channel are under manual control.
Procedure
Set Channel B as desired.
Option
Description
Frequency output
Channel B will operate as a frequency output.
Discrete output
Channel B will operate as a discrete output.
Integrate the meter with the control system
Configuration and Use Manual
99
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......