
Connection over network
Figure B-6:
A
C
E
D
B
A. PC
B. Signal converter
C.
120-
Ω
, 1/2-watt resistors at both ends of the segment, if necessary
D. DCS or PLC
E. Transmitter, with wiring compartment and power supply compartment opened
Note
This figure shows a serial port connection. USB connections are also supported.
4.
Start ProLink II.
5.
Choose
Connection
>
Connect to Device
.
6.
Set the connection parameters to the values configured in the transmitter.
If your transmitter has not been configured, use the default values shown here.
Default HART/RS-485 connection parameters
Table B-1:
Parameter
Default values
Protocol
HART
Baud
1200
Parity
Odd
Stop Bits
1
Address
0
Tip
If you do not know the transmitter’s RS-485 communication settings, you can connect
through the service port, which always uses default settings, or use another communications
tool to view or change the settings.
7.
Set the
COM Port
value to the PC COM port that you are using for this connection.
8.
Set
Master
as appropriate.
Using ProLink II with the transmitter
Configuration and Use Manual
283
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......