Status alarms and Status Alarm Severity
(continued)
Table 5-2:
Alarm code
Status message
Default severity Notes
Configurable?
A031
Low Power
Fault
Applies only to flowmeters with the
enhanced core processor.
No
A032
Meter Verification in Pro-
gress: Outputs to Fault
Varies
Applies only to transmitters with
Smart Meter Verification.
If outputs are set to
Last Measured
Value
, severity is
Info
. If outputs are
set to
Fault
, severity is
Fault
.
No
A033
Insufficient Right/Left Pick-
off Signal
Fault
Applies only to flowmeters with the
enhanced core processor.
Yes
A034
Meter Verification Failed
Informational
Applies only to transmitters with
Smart Meter Verification.
Yes
A035
Meter Verification Aborted
Informational
Applies only to transmitters with
Smart Meter Verification.
Yes
A100
mA Output 1 Saturated
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A101
mA Output 1 Fixed
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A102
Drive Overrange
Informational
Yes
A103
Data Loss Possible (Totals
and Inventories)
Informational
Applies only to flowmeters with the
standard core processor.
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A104
Calibration in Progress
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A105
Slug Flow
Informational
Yes
A106
Burst Mode Enabled
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A107
Power Reset Occurred
Informational
Normal transmitter behavior; oc-
curs after every power cycle.
Yes
A108
Basic Event 1 On
Informational
Applies only to basic events.
Yes
A109
Basic Event 2 On
Informational
Applies only to basic events.
Yes
A110
Frequency Output Satura-
ted
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A111
Frequency Output Fixed
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A112
Upgrade Transmitter Soft-
ware
Informational
Applies only to systems with trans-
mitter software earlier than v5.0.
Yes
A113
mA Output 2 Saturated
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
Configure device options and preferences
Configuration and Use Manual
93
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......