Effect of the Flow Direction parameter and actual flow direction on flow
values reported via digital communications
Table 4-8:
Flow Direction
setting
Actual flow direction
Forward
Zero flow
Reverse
Forward
Positive
0
Negative
Reverse
Positive
0
Negative
Bidirectional
Positive
0
Negative
Absolute Value
Positive
0
Positive
Negate Forward
Negative
0
Positive
Negate Bidirectional
Negative
0
Positive
Effect of Flow Direction on flow totals
Flow Direction
affects how flow totals and inventories are calculated.
Effect of the Flow Direction parameter and actual flow direction on flow
totals and inventories
Table 4-9:
Flow Direction
setting
Actual flow direction
Forward
Zero flow
Reverse
Forward
Totals increase
Totals do not change
Totals do not change
Reverse
Totals do not change
Totals do not change
Totals increase
Bidirectional
Totals increase
Totals do not change
Totals decrease
Absolute Value
Totals increase
Totals do not change
Totals increase
Negate Forward
Totals do not change
Totals do not change
Totals increase
Negate Bidirectional
Totals decrease
Totals do not change
Totals increase
4.5
Configure density measurement
The density measurement parameters control how density is measured and reported.
Density measurement (along with mass measurement) is used to determine liquid volume
flow.
The density measurement parameters include:
•
Density Measurement Unit
•
Slug Flow Parameters
•
Density Damping
•
Density Cutoff
(3) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
Configure process measurement
Configuration and Use Manual
45
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......