For D3 density calibration, the D3 fluid must meet the following requirements:
-
Minimum density of 0.6 g/cm
3
-
Minimum difference of 0.1 g/cm
3
between the density of the D3 fluid and the
density of water. The density of the D3 fluid may be either greater or less than
the density of water.
•
For D4 density calibration, the D4 fluid must meet the following requirements:
-
Minimum density of 0.6 g/cm
3
-
Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of the D3 fluid. The density of the D4 fluid must be greater than the
density of the D3 fluid.
-
Minimum difference of 0.1 g/cm
3
between the density of the D4 fluid and the
density of water. The density of the D4 fluid may be either greater or less than
the density of water.
•
Before performing the calibration, record your current calibration parameters. If the
calibration fails, restore the known values.
Procedure
See
Measurement support
202
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......