Status alarms and recommended actions
(continued)
Table 12-2:
Alarm code
Description
Recommended actions
A110
Frequency Output Saturated
The calculated frequency output is outside the configured
range.
1. Check the
Frequency Output Scaling Method
parameter.
2. Check your process conditions against the values reported
by the flowmeter.
3. Verify process conditions, checking especially for air in the
flow tubes, tubes not filled, foreign material in the tubes, or
coating in the tubes.
4. Verify that the measurement units are configured correctly
for your application.
5. Purge the flow tubes.
A111
Frequency Output Fixed
The frequency output has been configured to send a constant
value.
1. Stopping the totalizer will set the frequency output to zero.
Cycling power to the transmitter or restarting the totalizer
will restore the frequency output to normal operation.
2. Check whether the output is in loop test mode. If it is, unfix
the output.
3. Check whether the output has been set to a constant value
via digital communication.
A112
Upgrade Transmitter Software Contact Micro Motion.
A113
mA Output 2 Saturated
1. Check your process conditions against the values reported
by the flowmeter.
2. Verify process conditions, checking especially for air in the
flow tubes, tubes not filled, foreign material in the tubes, or
coating in the tubes.
3. Verify that the measurement units are configured correctly
for your application.
4. Purge the flow tubes.
5. Check the
Upper Range Value
and
Lower Range Value
parame-
ters. Are they set correctly?
A114
mA Output 2 Fixed
1. Check whether the output is in loop test mode. If it is, unfix
the output.
2. Exit mA output trim, if applicable.
3. Check whether the output has been set to a constant value
via digital communication.
A115
No External Input or Polled Da-
ta
The HART polling connection to an external device has failed. No
response received from polled device.
1. Verify the external device operation.
2. Verify the wiring between the transmitter and the external
device.
3. Verify the HART polling configuration.
A116
Temperature Overrange (Pe-
troleum)
1. Check your process conditions against the values reported
by the flowmeter.
2. Verify the configuration of the petroleum measurement ta-
ble type and temperature.
Troubleshooting
218
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......