Default values for Lower Range Value (LRV) and Upper Range Value (URV)
(continued)
Table 6-2:
Process variable
LRV
URV
Concentration
0%
100%
Baume
0
10
Specific gravity
0
10
6.2.3
Configure AO Cutoff
Display
Not available
ProLink II
ProLink
>
Configuration
>
Analog Output
>
Primary Output
>
AO Cutoff
ProLink III
Device Tools
>
Configuration
>
I/O
>
Outputs
>
mA Output
Field Communicator
Configure
>
Manual Setup
>
Inputs/Outputs
>
mA Output
>
mA Output Settings
>
MAO Cutoff
Overview
AO Cutoff
(Analog Output Cutoff) specifies the lowest mass flow rate, volume flow rate, or
gas standard volume flow rate that will be reported through the mA output. Any flow rates
below
AO Cutoff
will be reported as 0.
Restriction
AO Cutoff
is applied only if
mA Output Process Variable
is set to
Mass Flow Rate
,
Volume Flow Rate
, or
Gas
Standard Volume Flow Rate
. If
mA Output Process Variable
is set to a different process variable,
AO Cutoff
is
not configurable, and the transmitter does not implement the AO cutoff function.
Procedure
Set
AO Cutoff
as desired.
The default value for
AO Cutoff
is 0.0 g/sec.
Tip
For most applications, the default value of
AO Cutoff
should be used. Contact Micro Motion customer
service before changing
AO Cutoff
.
Interaction between AO Cutoff and process variable cutoffs
When
mA Output Process Variable
is set to a flow variable (for example, mass flow rate or
volume flow rate),
AO Cutoff
interacts with
Mass Flow Cutoff
or
Volume Flow Cutoff
. The
transmitter puts the cutoff into effect at the highest flow rate at which a cutoff is
applicable.
Integrate the meter with the control system
104
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......