1.
Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions. During the power-up sequence, Alarm A009 is active.
This alarm should clear automatically when the power-up sequence is complete.
2.
Check the status LED on the transmitter.
Transmitter status reported by status LED
Table 2-1:
LED state
Description
Recommendation
Green
No alarms are active.
Continue with configuration or process meas-
urement.
Flashing green
(1)
No alarms are active. One or more previously
active alarms have not been acknowledged.
Continue with configuration or process meas-
urement. If you choose, you can acknowledge
the alarms.
Yellow
One or more low-severity alarms are active,
and have been acknowledged.
A low-severity alarm condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or proc-
ess measurement. If you choose, you can iden-
tify and resolve the alarm condition.
Flashing yellow
(2)
One or more low-severity alarms are active
and have not been acknowledged.
A low-severity alarm condition does not affect
measurement accuracy or output behavior.
You can continue with configuration or proc-
ess measurement. If you choose, you can iden-
tify and resolve the alarm condition. You may
also acknowledge the alarm.
Red
One or more high-severity alarms are active,
and have been acknowledged.
A high-severity alarm condition affects meas-
urement accuracy and output behavior. Re-
solve the alarm condition before continuing.
Flashing red
(3)
One or more high-severity alarms are active
and have not been acknowledged.
A high-severity alarm condition affects meas-
urement accuracy and output behavior. Re-
solve the alarm condition before continuing.
You may also acknowledge the alarm.
Postrequisites
For information on viewing the list of active alarms, see
.
For information on individual alarms and suggested resolutions, see
(1) If
Status LED Blinking
is disabled, the LED will show solid green rather than flashing.
(2) If
Status LED Blinking
is disabled, the LED will show solid yellow rather than flashing.
(3) If
Status LED Blinking
is disabled, the LED will show solid red rather than flashing.
Quick start
8
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......