Effects of Update Rate =
Special
Incompatible features and functions
Special
mode is not compatible with the following features and functions:
•
Enhanced events. Use basic events instead.
•
All calibration procedures.
•
Zero verification.
•
Restoring the factory zero or the prior zero.
If required, you can switch to
Normal
mode, perform the desired procedures, and then
return to
Special
mode.
Process variable updates
Some process variables are not updated when
Special
mode is enabled.
Special
mode and process variable updates
Table 5-1:
Always polled and updated
Updated only when the petroleum
measurement application is disa-
bled
Never updated
•
Mass flow
•
Volume flow
•
Gas standard volume flow
•
Density
•
Temperature
•
Drive gain
•
LPO amplitude
•
Status [contains Event 1 and Event
2 (basic events)]
•
Raw tube frequency
•
Mass total
•
Volume total
•
Gas standard volume total
•
Temperature-corrected volume
total
•
Temperature-corrected density
•
Temperature-corrected volume
flow
•
Batch-weighted average tempera-
ture
•
Batch-weighted average density
•
RPO amplitude
•
Board temperature
•
Core input voltage
•
Mass inventory
•
Volume inventory
•
Gas standard volume inventory
All other process variables and calibra-
tion data. They retain the values held
at the time you enabled
Special
mode.
Configure device options and preferences
88
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......