a. Allow the flowmeter to warm up for at least 20 minutes after applying power.
b. Run the process fluid through the sensor until the sensor temperature reaches
the normal process operating temperature.
c. Stop flow through the sensor by shutting the downstream valve, and then the
upstream valve if available.
d. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
e. Observe the drive gain, temperature, and density readings. If they are stable,
check the
Live Zero
or
Field Verification Zero
value. If the average value is close to 0,
you should not need to zero the flowmeter.
2.
Choose
Device Tools
>
Calibration
>
Zero Verification and Calibration
.
3.
Click
Calibrate Zero
.
4.
Modify
Zero Time
, if desired.
Zero Time
controls the amount of time the transmitter takes to determine its zero-
flow reference point. The default
Zero Time
is 20 seconds. For most applications, the
default
Zero Time
is appropriate.
5.
Click
Calibrate Zero
.
The
Calibration in Progress
message is displayed. When the calibration is complete:
•
If the zero procedure was successful, a
Calibration Success
message and a new zero
value are displayed.
•
If the zero procedure failed, a
Calibration Failed
message is displayed.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Need help?
If the zero fails:
•
Ensure that there is no flow through the sensor, then retry.
•
Remove or reduce sources of electromechanical noise, then retry.
•
Set
Zero Time
to a lower value, then retry.
•
If the zero continues to fail, contact Micro Motion.
•
If you want to return the flowmeter to operation using a previous zero value:
-
To restore the zero value set at the factory:
Device Tools
>
Zero Verification and Calibration
>
Calibrate Zero
>
Restore Factory Zero
. This function requires the enhanced core processor.
-
To restore the most recent valid value from transmitter memory:
Device Tools
>
Zero
Verification and Calibration
>
Calibrate Zero
>
Restore Prior Zero
.
Restore Prior Zero
is available only
while the Flow Calibration window is open. If you close the Flow Calibration window, you
will no longer be able to restore the prior zero.
Restriction
Restore the factory zero only if your flowmeter was purchased as a unit, it was zeroed at the factory,
and you are using the original components.
Measurement support
190
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......