Flow measurement problems and recommended actions
(continued)
Table 12-3:
Problem
Possible causes
Recommended actions
Erratic non-zero flow
rate when flow is
steady
•
Slug flow
•
Damping value too low
•
Plugged or coated flow tube
•
Output wiring problem
•
Problem with receiving device
•
Wiring problem
•
Verify that the sensor orientation is appropri-
ate for your application (refer to the sensor
installation manual).
•
Check the drive gain and the pickoff voltage.
See
•
If the wiring between the sensor and the
transmitter includes a 9-wire segment, verify
that the 9-wire cable shields are correctly
grounded.
•
Check for air entrainment, tube fouling,
flashing, or tube damage.
•
Check the wiring between the sensor and
transmitter. See
.
•
For sensors with a junction box, check for
moisture in the junction box.
•
Purge the flow tubes.
•
Check for open or leaking valves or seals.
•
Check for sources of vibration.
•
Verify damping configuration.
•
Verify that the measurement units are con-
figured correctly for your application.
•
•
Check for radio frequency interference. See
.
•
Contact Micro Motion.
Inaccurate flow rate
or batch total
•
Wiring problem
•
Inappropriate measurement unit
•
Incorrect flow calibration factor
•
Incorrect meter factor
•
Incorrect density calibration factors
•
Incorrect flowmeter grounding
•
Slug flow
•
Problem with receiving device
•
Check the wiring between the sensor and
transmitter. See
.
•
Verify that the measurement units are con-
figured correctly for your application.
•
Verify that all of the characterization parame-
ters match the data on the sensor tag.
•
Perform a bucket test to verify batch totals.
•
Zero the meter.
•
•
•
Verify that the receiving device, and the wir-
ing between the transmitter and the receiv-
ing device.
•
Check sensor coil resistance and for shorts to
case. See
•
Replace the core processor or transmitter.
Troubleshooting
Configuration and Use Manual
221
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......