Derived variables and calculated process variables
Table 4-15:
Derived Variable
Description
Calculated process variables
Density at
reference
tempera-
ture
Standard
volume
flow rate
Specific
gravity
Concen-
tration
Net mass
flow rate
Net vol-
ume flow
rate
Density at reference
temperature
Mass/unit volume, cor-
rected to a given refer-
ence temperature
✓
✓
Specific gravity
The ratio of the density
of a process fluid at a
given temperature to
the density of water at
a given temperature.
The two given temper-
ature conditions do not
need to be the same.
✓
✓
✓
Mass concentration
derived from refer-
ence density
The percent mass of
solute or of material in
suspension in the total
solution, derived from
reference density
✓
✓
✓
✓
Mass concentration
derived from specif-
ic gravity
The percent mass of
solute or of material in
suspension in the total
solution, derived from
specific gravity
✓
✓
✓
✓
✓
Volume concentra-
tion derived from
reference density
The percent volume of
solute or of material in
suspension in the total
solution, derived from
reference density
✓
✓
✓
✓
Volume concentra-
tion derived from
specific gravity
The percent volume of
solute or of material in
suspension in the total
solution, derived from
specific gravity
✓
✓
✓
✓
✓
Concentration de-
rived from reference
density
The mass, volume,
weight, or number of
moles of solute or of
material in suspension
in proportion to the to-
tal solution, derived
from reference density
✓
✓
✓
Configure process measurement
68
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......