Postrequisites
•
To clear the following alarms, you must correct the problem, acknowledge the
alarm, then power-cycle the transmitter: A001, A002, A010, A011, A012, A013,
A018, A019, A022, A023, A024, A025, A028, A029, A031.
•
For all other alarms:
-
If the alarm is inactive when it is acknowledged, it will be removed from the list.
-
If the alarm is active when it is acknowledged, it will be removed from the list
when the alarm condition clears.
9.4.2
View and acknowledge alarms using ProLink II
You can view a list containing all alarms that are active, or inactive but unacknowledged.
From this list, you can acknowledge individual alarms.
1.
Choose
ProLink
>
Alarm Log
.
2.
Choose the High Priority or Low Priority panel.
Note
The grouping of alarms into these two categories is hard-coded and is not affected by
Status
Alarm Severity
.
All active or unacknowledged alarms are listed:
•
Red indicator: Alarm is currently active.
•
Green indicator: Alarm is inactive but unacknowledged.
Note
Only Fault and Informational alarms are listed. The transmitter automatically filters out alarms
with
Status Alarm Severity
set to
Ignore
.
3.
To acknowledge an alarm, check the
Ack
checkbox.
Postrequisites
•
To clear the following alarms, you must correct the problem, acknowledge the
alarm, then power-cycle the transmitter: A001, A002, A010, A011, A012, A013,
A018, A019, A022, A023, A024, A025, A028, A029, A031.
•
For all other alarms:
-
If the alarm is inactive when it is acknowledged, it will be removed from the list.
-
If the alarm is active when it is acknowledged, it will be removed from the list
when the alarm condition clears.
Transmitter operation
Configuration and Use Manual
155
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......