3.
Click
Calibrate Zero
.
4.
Modify
Zero Time
, if desired.
Zero Time
controls the amount of time the transmitter takes to determine its zero-
flow reference point. The default
Zero Time
is 20 seconds. For most applications, the
default
Zero Time
is appropriate.
5.
Click
Perform Auto Zero
.
The
Calibration in Progress
light will turn red during the zero procedure. At the end of
the procedure:
•
If the zero procedure was successful, the
Calibration in Progress
light returns to
green and a new zero value is displayed.
•
If the zero procedure failed, the
Calibration Failure
light turns red.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Need help?
If the zero fails:
•
Ensure that there is no flow through the sensor, then retry.
•
Remove or reduce sources of electromechanical noise, then retry.
•
Set
Zero Time
to a lower value, then retry.
•
If the zero continues to fail, contact Micro Motion.
•
If you want to return the flowmeter to operation using a previous zero value:
-
To restore the zero value set at the factory:
ProLink
>
Zero Verification and Calibration
>
Calibrate
Zero
>
Restore Factory Zero
.This function requires the enhanced core processor.
-
To restore the most recent valid value from transmitter memory:
ProLink
>
Zero Verification
and Calibration
>
Calibrate Zero
>
Restore Prior Zero
.
Restore Prior Zero
is available only while the
Flow Calibration window is open. If you close the Flow Calibration window, you will no
longer be able to restore the prior zero.
Restriction
Restore the factory zero only if your flowmeter was purchased as a unit, it was zeroed at the factory,
and you are using the original components.
11.3.3
Zero the flowmeter using ProLink III
Zeroing the flowmeter establishes a baseline for process measurement by analyzing the
sensor's output when there is no flow through the sensor tubes.
Prerequisites
ProLink III must be running and must be connected to the transmitter.
Procedure
1.
Prepare the flowmeter:
Measurement support
Configuration and Use Manual
189
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......