Options for HART variables
(continued)
Table 6-13:
Process variable
Primary Varia-
ble (PV)
Secondary
Variable (SV)
Third Variable
(TV)
Fourth Varia-
ble (QV )
ED standard volume flow rate
✓
✓
✓
✓
ED standard volume total
✓
ED standard volume inventory
✓
ED net mass flow rate
✓
✓
✓
✓
ED net mass total
✓
ED net mass inventory
✓
ED net volume flow rate
✓
✓
✓
✓
ED net volume total
✓
ED net volume inventory
✓
ED concentration
✓
✓
✓
ED Baume
✓
✓
✓
Interaction of HART variables and transmitter outputs
The HART variables are automatically reported through specific transmitter outputs. They
may also be reported through HART burst mode, if enabled on your transmitter.
HART variables and transmitter outputs
Table 6-14:
HART variable
Reported via
Comments
Primary Variable (PV)
Primary mA output
If one assignment is changed, the other is changed auto-
matically, and vice versa.
Secondary Variable (SV)
Not associated with an out-
put
The SV must be configured directly, and the value of the
SV is available only via digital communications.
Tertiary Variable (TV)
Frequency output (if
present on your transmitter)
If one assignment is changed, the other is changed auto-
matically, and vice versa. If your transmitter does not have
a frequency output, the TV must be configured directly,
and the value of the TV is available only via digital commu-
nications.
Quaternary Variable (QV)
Not associated with an out-
put
The QV must be configured directly, and the value of the
QV is available only via digital communications.
Integrate the meter with the control system
128
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......