If your sensor tag does not show a
DT
or
TC
value, enter the last 3 digits of the density
calibration factor. In the sample tag, this value is shown as 4.44 (see
Flow calibration parameters (FCF, FT)
Two separate values are used to describe flow calibration: a 6-character FCF value and a 4-
character
FT
value. Both values contain decimal points. During characterization, these may
be entered as two values or as a single 10-character string. The 10-character string is called
either
Flowcal
or
FCF
.
If your sensor tag shows the
FCF
and the
FT
values separately and you need to enter a
single value, concatenate the two values to form the single parameter value.
If your sensor tag shows a concatenated
Flowcal
or
FCF
value and you need to enter the
FCF
and the
FT
values separately, split the concatenated value:
•
FCF
= The first 6 characters, including the decimal point
•
FT
= The last 4 characters, including the decimal point
Example: Concatenating FCF and FT
FCF = x.xxxx
FT = y.yy
Flow calibration parameter: x.xxxxy.yy
Example: Splitting the concatenated Flowcal or FCF value
Flow calibration parameter: x.xxxxy.yy
FCF = x.xxxx
FT = y.yy
2.5
Verify mass flow measurement
Check to see that the mass flow rate reported by the transmitter is accurate. You can use
any available method.
•
Read the value for
Mass Flow Rate
on the transmitter display.
•
Connect to the transmitter with ProLink II and read the value for
Mass Flow Rate
in the
Process Variables window (
ProLink
>
Process Variables
).
•
Connect to the transmitter with ProLink III and read the value for
Mass Flow Rate
in
the Process Variables panel.
•
Connect to the transmitter with the Field Communicator and read the value for
Mass
Flow Rate
in the Process Variables menu (
On-Line Menu
>
Overview
>
Primary Purpose
Variables
).
Postrequisites
If the reported mass flow rate is not accurate:
•
Check the characterization parameters.
Quick start
Configuration and Use Manual
13
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......