3.
Set
Approval
to the appropriate regulatory agency for your application.
Option
Description
NTEP
Regulatory agency for the U.S.A. and Canada
OIML
Regulatory agency for all other world areas
4.
Set
Totalizer Reset Options
as desired.
Option
Description
Not Resettable from Display
and Digital Comm
When the transmitter is secured, totalizers cannot be reset
from either the transmitter display or digital communications.
Resettable from Digital Comm
Only
When the transmitter is secured, you must use digital commu-
nications to reset the totalizers.
Resettable from Display and
Digital Comm
When the transmitter is secured, you can use either the trans-
mitter display or digital communications to reset the totaliz-
ers.
Resettable from Display Only
When the transmitter is secured, you must use the transmitter
display to reset the totalizers.
“Digital communications” refers to any method that uses Modbus or HART
communications to interact with the transmitter. This includes ProLink II, ProLink III,
the Field Communicator, and any host.
5.
If required for your installation, configure your transmitters for two frequency
outputs and set them to operate in Quadrature mode.
a. Choose
ProLink
>
Configuration
>
Channel
.
b. Set
Channel B Type Assignment
to
Frequency Output
and click
Apply
.
c. Set
Channel C Type Assignment
to
Frequency Output
and click
Apply
.
d. Choose
ProLink
>
Configuration
>
Frequency
and set
Freq Output Mode
to
Quadrature
.
6.
If required for your installation, enable
Alarm Menu Password
for the display:
ProLink
>
Configuration
>
Display
.
Alarm Menu Password
is required under German law for PTB-type approval for gas
applications, and may be required in other locations or for other application types.
7.
If required for your installation, read and record the firmware checksum values for
the transmitter and the core processor:
ProLink
>
Configuration
>
Device
.
The checksum values for the transmitter firmware and the core processor firmware
must be read during meter commissioning to meet Weights & Measures
requirements for gas applications in Germany. They may also be useful for
MID/Welmec 7.2 test reports.
8.
If required for your installation, configure
Field Verification Zero
(
FVZ
) as a display
variable, then observe and record the current value of
FVZ
:
ProLink
>
Configuration
>
Display
.
Set up the Weights & Measures application
Configuration and Use Manual
141
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......