13.
Set
Active Matrix
to the matrix to be used for measurement.
Concentration process variables are now available on the transmitter. You can view and
report them in the same way that you view and report other process variables.
4.8.3
Configure concentration measurement using the
Field Communicator
This task guides you through setting up a concentration matrix to use for measurement. It
does not cover loading or building a concentration matrix.
Note
Concentration matrices can be made available on your transmitter either by loading an existing
matrix from a file or by building a new matrix. Up to six matrices can be available on your transmitter,
but only one can be used for measurement at any given time. See
Micro Motion Enhanced Density
Application: Theory, Configuration, and Use
for detailed information on building a matrix.
Prerequisites
Before you can configure concentration measurement:
•
The concentration measurement application must be enabled on your transmitter.
•
You must know the derived variable that your matrix is designed for.
•
You must know the density unit used by your matrix.
•
You must know the temperature unit used by your matrix.
•
The concentration measurement application must be unlocked.
Procedure
1.
Choose
Online
>
Configure
>
Manual Setup
>
Measurements
>
Density
and set
Density Unit
to
match the density unit used by your matrix.
2.
Choose
Online
>
Configure
>
Manual Setup
>
Measurements
>
Temperature
and set
Temperature Unit
to match the temperature unit used by your matrix.
3.
Choose
Online
>
Configure
>
Manual Setup
>
Measurements
>
Conc Measurement (CM)
>
CM
Configuration
.
4.
Set up extrapolation alerts.
Each concentration matrix is built for a specific density range and a specific
temperature range. If process density or process temperature goes outside the
range, the transmitter will extrapolate concentration values. However, extrapolation
may affect accuracy. Extrapolation alerts are used to notify the operator that
extrapolation is occurring.
a. Choose
Online
>
Configure
>
Manual Setup
>
Measurements
>
Conc Measurement (CM)
>
Matrix Configuration
.
b. Set
Matrix Being Configured
to the matrix that you want to configure.
c. Set
Extrapolation Alert Limit
to the point, in percent, at which an extrapolation alert
will be posted.
Configure process measurement
64
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......