Option
Setup
A value written by
digital communica-
tions
a. Choose
Online
>
Configure
>
Manual Setup
>
Measurements
>
External
Pressure/Temperature
>
Temperature
.
b. Set
External Temperature
to
Enabled
.
c. Perform the necessary host programming and communications
setup to write temperature data to the transmitter at appropri-
ate intervals.
Note
If the Weights & Measures application is implemented and the trans-
mitter is secured, digital communications cannot be used to write
temperature data to the transmitter.
4.7.4
API reference tables
API reference tables, associated process fluids, and associated calculation values
Table 4-13:
Table
name
Process fluid
CTL source data
Reference temperature Density unit
5A
Generalized crude and
JP4
Observed density and
observed temperature
60 °F (non-configurable) Degrees API
Range: 0 to 100
5B
Generalized products
Observed density and
observed temperature
60 °F (non-configurable) Degrees API
Range: 0 to 85
5D
Lubricating oils
Observed density and
observed temperature
60 °F (non-configurable) Degrees API
Range:
−
10 to +40
6C
Liquids with a constant
density base or known
thermal expansion coef-
ficient
User-supplied reference
density (or thermal ex-
pansion coefficient) and
observed temperature
60 °F (non-configurable) Degrees API
23A
Generalized crude and
JP4
Observed density and
observed temperature
60 °F (non-configurable) Relative density
Range: 0.6110 to 1.0760
23B
Generalized products
Observed density and
observed temperature
60 °F (non-configurable) Relative density
Range: 0.6535 to 1.0760
23D
Lubricating oils
Observed density and
observed temperature
60 °F (non-configurable) Relative density
Range: 0.8520 to 1.1640
24C
Liquids with a constant
density base or known
thermal expansion coef-
ficient
User-supplied reference
density (or thermal ex-
pansion coefficient) and
observed temperature
60 °F (non-configurable) Relative density
53A
Generalized crude and
JP4
Observed density and
observed temperature
15 °C (configurable)
Base density
Range: 610 to
1075 kg/m
3
Configure process measurement
Configuration and Use Manual
57
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......