The flow factor is the percent change in the flow rate per PSI. When entering the
value, reverse the sign.
Example:
If the flow factor is 0.000004 % per PSI, enter
−
0.000004
% per PSI.
4.
Enter
Density Factor
for your sensor.
The density factor is the change in fluid density, in g/cm
3
/PSI. When entering the
value, reverse the sign.
Example:
If the density factor is 0.000006 g/cm
3
/PSI, enter
−
0.000006
g/cm3/PSI.
5.
Enter
Cal Pressure
for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and
defines the pressure at which there is no pressure effect. If the data is unavailable,
enter
20
PSI.
6.
Determine how the transmitter will obtain pressure data, and perform the required
setup.
Option
Setup
A user-configured
static pressure val-
ue
a. Set
Pressure Units
to the desired unit.
b. Set
External Pressure
to the desired value.
Polling for pressure
a. Ensure that the primary mA output has been wired to support
HART polling.
b. Choose
ProLink
>
Configuration
>
Polled Variables
.
c. Choose an unused polling slot.
d. Set
Polling Control
to
Poll As Primary
or
Poll as Secondary
, and click
Apply
.
e. Set
External Tag
to the HART tag of the external pressure device.
f. Set
Variable Type
to
Pressure
.
Tip
•
Poll as Primary
: No other HART masters will be on the network.
•
Poll as Secondary
: Other HART masters will be on the network.
The Field Communicator is not a HART master.
Configure process measurement
70
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......