Option
Description
Enabled
(default)
Operators can use a single display command to acknowledge all alarms
at once.
Disabled
Operators cannot acknowledge all alarms at once, they must be ac-
knowledged individually.
5.3
Configure security for the display menus
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
DISPLAY
ProLink II
ProLink
>
Configuration
>
Display
>
Display Options
ProLink III
Device Tools
>
Configuration
>
Transmitter Display
>
Display Security
Field Communicator
Configure
>
Manual Setup
>
Display
>
Offline Variable Menu Features
Overview
You can control operator access to different sections of the display off-line menu. You can
also configure a password to control access.
Procedure
1.
To control operator access to the maintenance section of the off-line menu, enable
or disable
Off-Line Menu
.
Option
Description
Enabled
(de-
fault)
Operator can access the maintenance section of the off-line menu. This
access is required for configuration and calibration, but is not required to
view alarms or to access Smart Meter Verification (if applicable).
Disabled
Operator cannot access the maintenance section of the off-line menu.
2.
To control operator access to the alarm menu, enable or disable
Alarm Menu
.
Option
Description
Enabled
(default)
Operator can access the alarm menu. This access is required to view and
acknowledge alarms, but is not required for Smart Meter Verification (if
applicable), configuration, or calibration.
Disabled
Operator cannot access the alarm menu.
Note
The transmitter status LED changes color to indicate that there are active alarms, but does
not show specific alarms.
Configure device options and preferences
Configuration and Use Manual
85
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......