Procedure
Set
Standard Gas Density
to the standard reference density of the gas you are measuring.
Note
ProLink II and ProLink III provide a guided method that you can use to calculate the standard density
of your gas, if you do not know it.
4.3.3
Configure Gas Standard Volume Flow Measurement Unit
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
UNITS
>
GSV
ProLink II
ProLink
>
Configuration
>
Flow
>
Std Gas Vol Flow Units
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
Field Communicator
Configure
>
Manual Setup
>
Measurements
>
GSV
>
GSV Flow Unit
Overview
Gas Standard Volume Flow Measurement Unit
specifies the unit of measure that will be displayed
for the gas standard volume flow rate. The measurement unit used for the gas standard
volume total and the gas standard volume inventory is derived from this unit.
Prerequisites
Before you configure
Gas Standard Volume Flow Measurement Unit
, be sure that
Volume Flow Type
is set to
Gas Standard Volume
.
Procedure
Set
Gas Standard Volume Flow Measurement Unit
to the unit you want to use.
The default setting for
Gas Standard Volume Flow Measurement Unit
is
SCFM
(Standard Cubic
Feet per Minute).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Gas Standard Volume Flow Measurement Unit
The transmitter provides a standard set of measurement units for
Gas Standard Volume Flow
Measurement Unit
, plus one user-defined special measurement unit. Different
communications tools may use different labels for the units.
Configure process measurement
36
Micro Motion
®
Model 2700 Transmitters with Analog Outputs
Summary of Contents for MICRO MOTION 2700
Page 10: ...Getting Started 2 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 14: ...Before you begin 6 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 26: ...Configuration and commissioning 18 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 84: ...Configure process measurement 76 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 214: ...Measurement support 206 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 258: ...Troubleshooting 250 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 354: ...Default values and ranges 346 Micro Motion Model 2700 Transmitters with Analog Outputs ...
Page 373: ......