PowerVault ML6000 Maintenance Guide
419
2
If the issue is a specific EKM server configuration problem, a separate RAS ticket has been generated
and requires resolution. Close this ticket and resolve the respective EKM server configuration ticket(s).
3
If the issue is a communication problem, perform the following steps:
a. Ensure that the library is properly connected to the network, and that the network cable is not
damaged.
b. Ensure that the library's IP address settings for the EKM servers are correct, and that the Key
Server IP and port configurations are configured accordingly.
c. Verify that the EKM servers are powered on, operational, and properly connected to the
network.
d. Verify that the Key Servers' IP and port configurations are properly set and responding to “ping”
requests.
e. If no problems are found with any of these items, there is a configuration problem with your
network. Contact your network administrator for assistance with troubleshooting this issue.
4
If the issue cannot be resolved, contact Service for further assistance.
5
You have three options:
a. Select
Close
to close the ticket now.
b. Select
Exit
to leave the ticket open for future troubleshooting.
c. Select
Back
to return to the Diagnostic Resolution.
DR137 - Resolve Media Threshold Notification
This DR provides customer resolution steps for
T147
on page 346
and
T148
on page 347.
Problem: The library has been configured to report when a tape cartridge capacity and/or tape cartridge load
count threshold is exceeded. A threshold has been met or exceeded.
1
Note the ticket details to determine which tape cartridge exceeded the media capacity or load count
threshold.
2
Depending on which threshold was exceeded, initiate any action that should be performed upon such
threshold notification warning.
3
You have three options:
a. Select
Close
to close the ticket now.
b. Select
Exit
to leave the ticket open for future troubleshooting.
c. Select
Back
to return to the Diagnostic Resolution.
DR146 - Resolve Drive Tape Alert 54 - No Start of Data
This DR provides customer resolution steps for
T188
on page 360.
Problem: A drive reported an issue during the start of data detection. This problem may be spurious, and
the drive may have recovered and now be reading the tape just fine; or the this problem may require further
analysis, as it may be caused by the specific tape cartridge, drive, or interaction of the media and drive.
1
View the ticket details and note the tape cartridge, drive type, and location, as well as the drive error
code.
2
If the drive error code lists error 7122 for a drive, the Format Identification Dataset (FID) is not properly
recorded, and data cannot be written to tape. The tape cartridge does not contain data and needs to be
discarded/replaced.
Summary of Contents for PowerVault ML6000
Page 1: ...w w w d e l l c o m s u p p o r t d e l l c o m MaintenanceGuide Dell PowerVault ML6000 ...
Page 14: ...xiv Tables ...
Page 18: ...xviii Figures ...
Page 452: ...434 Troubleshooting the Tape Library ...
Page 456: ...438 Library Drawings Library Rear View Figure 32 ML6020 Rear View ...
Page 468: ...450 Library Drawings Power Supplies Figure 49 Power Supplies ...
Page 477: ...PowerVault ML6000 Maintenance Guide 459 Figure 58 Electrical System 5U Library Control Module ...
Page 478: ...460 Library Drawings Figure 59 LCB Functions ...
Page 479: ...PowerVault ML6000 Maintenance Guide 461 ...
Page 482: ...464 Library Drawings Figure 64 Fan Control Board ...
Page 483: ...PowerVault ML6000 Maintenance Guide 465 Figure 65 Library Cables ...
Page 484: ...466 Library Drawings Figure 66 Library System Grounding ...
Page 485: ...PowerVault ML6000 Maintenance Guide 467 Figure 67 Library Power System ...
Page 486: ...468 Library Drawings ...
Page 494: ...476 Glossary Y axis Y position The vertical position of the library s robotic arm ...