137
Instruction Manual
Table A4-3.
Harvest set up check list.
Label
Harvest Set Up Check List
Check
Reference
Procedure
Scan bar code on Harvest kit sterile bag.
Section 8.4.1
Fill the Harvest Wash bottle with DPBS and the Harvest Solution bottle with e.g., Accutase® in an aseptic
environment.
1
Set up blue tubing lines in blue pinch valve.
Tip:
Install the white tubing line first then the black tubing line.
2
Set up green tubing lines in green pinch valve.
Tip:
Aseptically connect the white/green tubing line to the Harvest Collection bottle.
3
Set up yellow tubing lines in yellow pinch valve.
4
Connect the Flush bottle air line to the Harvest pressure port.
5
Select Harvest Preparation on the HMI.
NOTE:
All equipment module controls, heaters, and pumps are turned off.
6
Clamp the bioreactor inlet and return tubing lines to the MCV.
7
Load yellow tubing line in the recirculation pump.
8
Connect the Harvest kit to the bioreactor inlet and outlet. Mate AseptiQuik® connectors to an audible click.
9
Remove both clamps on the Harvest lines.
Harvest Collection
Table A4-4.
Harvest collection check list.
Label
Harvest collection check list
Check
Reference
Procedure
1
Start Harvest Wash Phase.
Section 8.4.2
2
Select Cell Release phase at user-specified duration.
Section 8.4.3
3
Start high pressure Cell Removal Phase for collection. Repeat Cell Release and Cell Removal phases, if
necessary.
Section 8.4.4
4
Select Final Rinse phase. Repeat as needed.
Section 8.4.5
5
Start the Cell Removal phase to collect the washed solution in the Harvest Collection bottle. Repeat Final
Rinse and Cell Removal phases as necessary.
Section 8.4.4
6
Select Shutdown to generate batch process file.
Section 8.6