85
Instruction Manual
3.
Use the positive pressure displacement bulb to feed the solution into the MCV.
The system begins auto-sequence steps. Return to Section 8.2.5 to sample the media.
8.2.5.2 FBR Inoculation
While FBR inoculation is a programmed option, MCV inoculation is the preferred method.
1. FBRX Inoculation Option.
Select FBRX Inoculation.
Press
to confirm.
Enter setpoints:
• Air MFC Flow Rate
• Stabilization Delay
Press
to confirm.
2. Connect Inoculation bottle, open and close respective clamps.
• Connect Inoculation bottle to the FBR inoculation port.
• Open clamp on the tubing line to the FBR Inoculation port.
• Close the clamp on the recirculation line.
Return media line
Return media line
Recirculation line
FBR Inoculation port
Return media line
Return media line
FBR Inoculation port
FBR Inoculation line clamp
(open)
Recirculation line
Return media line
Return media line
FBR Inoculation port
Recirculation line
Recirculation line clamp (close)