65
Instruction Manual
Protocols that ensure system integrity are found in Table 8-2.
Table 8-2.
Calibration protocols.
Protocol
Reference
Pump Calibration
Section 8.1
Sensor Calibration
Section 8.2.3.1
Reminder 1:
If the green button with an exclamation mark is flashing green, press to view, and address the task in the prompt window.
Press
after the manual task is completed to resume (Figure 8-2).
Reminder 2:
The system will remain in its current phase or state until the user acknowledges step completion. Phases do not start
automatically. Only some sub-steps in some phases may start automatically.
Figure 8-2.
Press the flashing green status box to
view outstanding tasks.
Figure 8-3.
Secure AseptiQuik® connection to an
audible click.
Figure 8-4.
Pull out both tabs after AseptiQuik®
connection is made.
Below are tips for running a seamless experiment:
Tip 1:
Make sure the correct bioreactor size is selected.
Tip 2:
Tightly mate AseptiQuik® connections to an audible click (Figure 8-3).
Tip 3:
Pull out to remove the blue tabs after mating the AseptiQuik® connections (Figure 8-4).
Tip 4:
Clamp any unused lines with AseptiQuik® connectors (Figure 8-5). This prevents liquid leaks when pressure is applied or if air is
drawn inside the tubing.
Tip 5:
Pump tubing must be centered in the pump head V-groove clamp (Figure 8-6). Improper tubing placement can cause pump failures.