This illustration shows an example of MCV media flow to the Lynx® connector before perfusing
the bioreactor.
8. Media perfusion.
• This illustration shows the bioreactor perfused with media (from the inlet at the base).
• Media flow continues up along the bioreactor outlet to the return media tubing.
9. Return media
This illustration shows media flow from the MCV to the FBR and returning to the MCV.
10. Media prime updates.
• The MCV level will update after priming.
• System Volume = MCV BRX Loop Volume.
11.
Press
to confirm Media Fill/ Prime phase is completed.
8.2.3 Media Conditioning Phase
This procedure shows how to stabilize temperature, DO, and pH for Media Conditioning. During this phase, probe calibrations are
performed. It is recommended that the system in Media Conditioning Phase is in cycle for a minimum of 2 hours, or preferably overnight.
1.
Select Media Phases.
Select Media Conditioning.