67
Instruction Manual
4.
Prime the calibration tubing with water first.
Press the appropriate pump icon on the HMI.
5.
A pop-up faceplate will appear. Follow the steps below:
1. Press Operator. A second pop-up screen will appear.
2. Press the Operator icon.
3. Enter Setpoint (mL/min.)
4. Jog the pump to fill the line with calibration solution. Jog until all air has been removed from
the calibration tubing and replaced with DI water.
Tip:
To activate a pump jog, tap and hold the jog button.
6.
Place the calibration tubing outlet end into an appropriately sized graduated cylinder or bottle.
This example shows a calibration tubing loaded in the media pump.
7.
Enter Pump Speed, and Seq. Duration.
1. Select Sample Type (Mass or Volume).
2. Press Start.
8.
The timer will start running during calibration.