Instruction Manual
2. Connect the media addition line.
• Flip open the AseptiQuik® connector blue tab on the media addition line and media addition
bottle line.
• Follow procedure in Section 8.2.1.2.
• Release the white plastic clamp on the media addition line so that it is open.
• Keep unused AseptiQuik® connector lines clamped.
The media addition bottle is aseptically connected and ready for batch and prime.
On the HMI, select Media Phases. Select Media Fill/Prime.
3. Media fill.
Enter setpoints:
• Media Flow Setpoint
• Media Fill Volume
• Recirculation Pump Flow Setpoint
Press
to start Media Fill.
NOTE:
The FBR size is automatically determined from the FBR bar code reading and will be
stored in the BRX Loop Volume under the Data button.
• Media from the media addition bottle is pumped towards the MCV.
• Check MCV liquid increase using the printed graduations on the vessel.