97
Instruction Manual
2.
Select Auxiliary Phases.
Select Media Addition.
3.
Enter setpoints:
- Media Flow Setpoint
- Media Addition Volume
- Number of Cycles
- Media Addition OFF Duration
• Press
to run Media Addition Phase.
• Media addition can be done as a single cycle (all at once) or in multiple cycles depending on
user’s protocol.
NOTE:
The system automatically limits Media Addition Volume to prevent sending a volume to
the MCV which exceeds the fill limit. The window with flashing red and white fields warns the
user of the error.
4.
Press
to acknowledge prompt that Media Addition is complete.
8.3.4 Bolus Addition
This phase is used to add feeds/bolus to the MCV at user-defined instances.
NOTE:
The user is responsible for visually monitoring the media level in the MCV to ensure levels are not unintentionally rising above or
sinking below the expected level over time.
1.
Select Auxiliary Phases.
Select Bolus Addition.