LeveLine
®
Instruction Manual
10401-01026 Rev D
General
LeveLine - CTD
LeveLine Mini - CTD
Temperature
ranges
Operational: -20 - 80° C (-4 - 176° F)
Storage: -40-80° C (-40-176° F)
Non Freezing
Operational: -20 - 80° C (-4 - 176° F)
Storage: -40-80° C (-40-176° F)
Non Freezing
Diameter
22mm (0.785 in)
22mm (0.785 in)
Length
245mm (9.65 in)
177mm (6.97 in)
Weight
180g (0.40 lb)
155g (0.34lb)
Materials
Titanium body, Delrin nose cone & hanger
Titanium body, Delrin nose cone & hanger
Output options
Modbus/RS485,
SDI-12,
Aquaread proprietary
Modbus/RS485,
SDI-12,
Aquaread proprietary
Battery type &
typical life
3.6V lithium, up to 10 years dependent on log
rate
N/A
External power
5 - 24 VDC
5 - 24 VDC
Memory Size
8.0 MB
N/A
Data record storage
500,000
N/A
Log types
Linear, Event & User-Selectable Schedule with
Future Start, Future Stop, Deploy Start and Real
Time View
N/A
Fastest logging rate
1 per second
N/A
Fastest Modbus
output rate
1 per second
1 per second
Fastest SDI-12
output rate
1 per second
1 per second
Real-time clock
(note 2)
Accurate to ± 2 second/24-hr period (± 15
minutes/year)
N/A
Pressure Sensor
Type / Material
Piezoresistive: ceramic
Piezoresistive: ceramic
Range (Absolute)
10M (33 ft)
20M (65 ft)
50M (164ft)
100M (3268ft)
10M (33 ft)
20M (65 ft)
50M (164ft)
100M (3268ft)
Range (Gauge)
10M (33 ft)
20M (65 ft)
50M (164ft)
10M (33 ft)
20M (65 ft)
50M (164ft)
Maximum pressure
Max 2x range, Burst 2.5x range
Max 2x range, Burst 2.5x range
Accuracy @ 25° C
(note 1)
±0.05% FS
±0.05% FS
Accuracy (overall)
(note 2)
±0.1% FS
±0.1% FS
Resolution
0.002% FS or better
0.002% FS or better
Electrical Conductivity
Range
0 – 200,000µS/cm
0 – 200,000µS/cm
Resolution
1µS/cm
1µS/cm
Accuracy
+/-1% reading or +/- 1µS/cm
whichever is greater See note 3
+/-1% reading or +/- 1µS/cm
whichever is greater See note 4
Salinity (see note 4)
Range
0 – 70PSU (ppt)
0 – 70PSU (ppt)
Resolution
0.01PSU (ppt)
0.01PSU (ppt)
Accuracy
+/-1% reading or +/- 0.1PSU
whichever is greater See note 3
+/-1% reading or +/- 0.1PSU
whichever is greater See note 3
Temperature Sensor
Accuracy
±0.1° C
±0.1° C
Resolution
0.01° C
0.01° C
Warranty
Standard
2 years
2 years
Notes:
1.
Across factory calibrated pressure range at constant temperature.
2.
Across factory calibrated pressure and temperature range (5°C - 25°C). Accuracy will degrade outside
these ranges.
3.
At the calibration point at 25°C.
4.
Reading calculated from EC and temperature values.
5.
Aquaread
®
Ltd reserves the right to change specifications without notice.
© 2019 Aquaread
®
Ltd.
www.aquaread.com
Page 90 of 90