LeveLine
®
Instruction Manual
10401-01026 Rev D
3.3. Deploying the Baro Logger
•
Measure the distance from the proposed suspension point to the estimated highest
surface level of the water.
•
Cut a length of cable to a length that will ensure the Baro logger will always remain
above
the maximum water level.
•
Secure the cable to the logger cap and the suspension point, preferably using a
swaged stainless steel ferrule and a carabiner clip as shown above.
•
If the Baro logger has not been programmed to start logging at some time in the
future, start the Baro logger now using either a LeveLine
®
QuickDeploy
™
Key, a
LeveLine
®
Meter or a PC running LeveLink
™
software.
•
When you are certain logging has started, screw the Baro logger firmly into the
logger cap and lower it into the mounting tube, ensuring it is well above the water
level.
3.4.
Using the LeveLine
®
QuickDeploy™ Key
The QuickDeploy
™
Key is a simple plug-in device which allows a pre-programmed logging
scheme to be started at the exact instant of deployment. It also embeds a zero depth
reference marker in the logged data in order to allow absolute depth measurement from the
start and carries out an instant check on the battery and memory levels.
3.4.1. Setup Prior To Deployment with a QuickDeploy™ Key
If you intend to make use of a QuickDeploy
™
Key, it is essential to set the LeveLine
®
Logger up first by connecting it to a PC running LeveLink
™
software.
Using the LeveLink
™
software, set the Logger's clock, site
ident, deployment position, logging rate, logging duration
and any event triggering you want.
Next, set the Manual Deployment option as shown here.
The Logger is now ready for deployment using a
QuickDeploy
™
Key.
If the Logger has been previously setup with a scheduled start, which is still in the future,
using the QuickDeploy
™
Key will override the scheduled start.
3.4.2. Deploying the LeveLine
®
Using the QuickDeploy™ Key
When you have completed your suspension cable installation as described in section
align the LED on the QuickDeploy
™
Key with the alignment spot on the LeveLine
®
Logger
and plug it in as shown in the photograph.
The LED on the QuickDeploy
™
Key will immediately flash to indicate the Logger's status in
accordance with the table below.
© 2019 Aquaread
®
Ltd.
www.aquaread.com
Page 44 of 90