2.8.7. LeveLine
®
Manual Logging Start, Logging Interval and Duration Setup
The LeveLine
®
Logger can be set up to start logging data automatically at a pre-determined
time for a set duration. This has been fully covered in the previous section.
If, on the other hand, you are not certain of the exact time and date you want to start
logging, the Logger can be set up for manual deployment. This involves setting up the
logging interval and optional logging duration, then forcing the logger to start logging at the
exact time of deployment using a PC running LeveLink
™
software, a LeveLine
®
Meter or a
QuickDeploy
™
Key.
First, set the logging interval as described in the previous section.
Next, select Manual Deployment as shown above. Finally, select the number of hours you
want to log data for.
Since, in this case, you are not specifying a time and date for logging to start, the End Time
and End Date boxes will remain at 'Not Set'. The actual time and date the logged data will
end will be the chosen number of hours after you manually deploy the Logger.
If you want to log data indefinitely after the manual deployment, set the value in the 'Log
data for:' to zero.
Finally, once the logging interval and duration (if required) have been set up, transfer the
settings to the Logger by clicking the 'Update Logging' button. You will notice that the
'Logging Status' will now show 'Not Scheduled' See illustration below.
Remember to subsequently start the LeveLine
®
logging when you deploy it!