LeveLine
®
Instruction Manual
10401-01026 Rev D
4. LeveLine
®
Meter
The LeveLine
®
Meter has four main functions.
•
Displaying live data from any model of LeveLine
®
Logger.
•
Allowing on-site setup and deployment of any model of LeveLine
®
Logger.
•
Allowing on-site, in-situ retrieval of data from up to 300 separate LeveLine
®
Loggers
for later download to a PC running LeveLink
™
software.
•
Calibration of the EC function on -CTD version Probes.
When being used in conjunction with a Meter, all LeveLine
®
Logging Probes are referred to
as 'Probes' rather then 'Loggers' in order to retain the naming conventions seen in other
models of Aquaread
®
Meters.
4.1. The LeveLine
®
Meter and the Environment
The LeveLine
®
Meter is designed to be used outdoors and is rated to IP67, that is to say it
is waterproof but it
is not
designed for submersion. In order to prevent accidental dunking
or loss, a lanyard is supplied.
Please note that the socket on the LeveLine
®
Meter is only waterproof when the
associated plug is fitted. Without the plug fitted, water can enter the socket. Damage
caused by water ingress through the socket is not covered by your warranty.
You may notice a small hole on the rear of the unit near the top. This is a waterproof vent
for the internal barometric sensor.
Do not poke anything in this hole!
Doing so will cause
major damage to the vent’s waterproof membrane and invalidate your warranty.
4.2. Meter Battery Installation and Care
The LeveLine
®
Meter requires five AA size batteries. To install the batteries, loosen the two
screws on the centreline of the rear of the meter and remove the battery compartment lid.
Following the battery polarity markings inside the battery compartment, insert five AA cells
then replace the compartment lid and tighten the screws.
4.2.1. Choice of Battery Type
Alkaline or rechargeable batteries may be used, but never mix battery types in the meter. If
you choose to use rechargeable batteries, we recommend
Energizer
2500mAh (or greater)
Nickel-Metal Hydride cells, which are widely available.
If the Meter is to be out of use for a long period, remove the batteries to prevent damage
due to possible leakage.
4.2.2. Battery Life
A set of fresh alkaline cells will give over 25 hours continuous use in the LeveLine
®
Meter.
A fully charged set of 2500mAh NiMH cells will give up to 50 hours continuous use.
4.2.3. Battery Charging
During the charging process, batteries generate heat and vent gasses, and must never be
charged inside a sealed unit. Because the LeveLine
®
Meter is a sealed unit, we do not
allow charging in-situ. Batteries must be removed and charged with a suitable battery
charger outside the Meter. We recommend the use of one of the
Energizer
range of NiMH
chargers.
© 2019 Aquaread
®
Ltd.
www.aquaread.com
Page 46 of 90