The desiccant used within the Aquaread® Desiccant Cartridges is a non-toxic, self-
indicating synthetic amorphous silica designed to change colour as it absorbs moisture
from orange to clear. Desiccant Cartridges are available in two sizes.
The choice of small (DC-50) or large (DC-100) Desiccant Cartridge is dependant upon the
humidity at the deployment site and the expected maintenance frequency. If the
deployment site is humid or the time between maintenance visits is expected to be long, the
large (DC-100) Desiccant Cartridge is recommended.
The time taken for a fresh, dry Desiccant Cartridge to go from orange to clear (i.e. become
fully saturated) is dependant upon site conditions. Regular maintenance checks should be
made on the installation and the Desiccant Cartridge should be replaced or regenerated
before it is completely saturated (clear).
To fit an Extension Cable Desiccant Housing (Part No. EX-DH), simply screw the free end
of the Extension Cable onto the Housing as shown in the photograph below. Once fitted,
remove the cap from your chosen Desiccant Cartridge and screw the Cartridge into the
Housing. Do not over-tighten the Cartridge.
The two vent holes located in the Desiccant Housing allow the pressure in the logger and
cable to remain at ambient atmospheric. Air is allowed to vent slowly in and out of the
assembly via an internal waterproof membrane. Any moisture that develops within the
cable or logger due to condensation is absorbed by the Desiccant Cartridge.
Do not block
the vent holes or allow the housing to sit in water.