LeveLine
®
Instruction Manual
10401-01026 Rev D
The LeveLine
®
Loggers are designed to connect to a LeveLine
®
Meter or a third party
telemetry device using a LeveLine
®
Extension Cable. The LeveLine
®
Extension Cable
features high-pressure metal connectors, which incorporate several O-ring seals at the
Logger end. Prior to first connection, the seals must be lubricated using the silicone grease
provided.
Apply a generous smear of grease to the connector O-rings situated just inside the
moveable retaining collar on the plug. Be careful not to get any grease inside the connector
near the gold contacts. A small smear of grease should also be applied to the thread on the
Logger to allow easy tightening of the collar.
To connect the Extension Cable to the LeveLine
®
Logger, align the spot on the LeveLine
®
Logger with the
◄AQUAREAD
logo on the plug body, then press the plug into the Logger
and tighten the retaining collar fully.
DO NOT TWIST THE CONNECTOR BODY WITH
RESPECT TO THE LOGGER
. Once the LeveLine
®
has been connected to the Extension
Cable, the LeveLine
®
Meter or telemetry device can be connected along with a desiccant
housing and desiccant cartridge.
3.2.6. Fitting a Desiccant Housing and Cartridge
When installing a gauge (vented) type LeveLine
®
logger with a vented cable,
the free end
of the cable must be positioned above the maximum expected water level.
It is also
essential to fit a desiccant housing and cartridge to the free end of the cable in order to
prevent any moisture entering the cable or logger.
If you intend to use a telemetry device with a permanent connection to the cable, a
Telemetry Cable Desiccant Housing (Part No. EX-TDH shown below) is required along with
either a DC-50 or DC-100 Desiccant Cartridge.
If the end of the cable is to be used only as a vent for the logger, an Extension Cable
Desiccant Housing (Part No. EX-DH) is required along with either a DC-50 or DC-100
Desiccant Cartridge. See the photograph at the bottom of the next page.