LeveLine
®
Instruction Manual
10401-01026 Rev D
The input registers provide a way to read the LeveLine's measurements and are
summarised in the tables below.
Register
Number
Register
Address (used
in Modbus
commands)
Description: Standard LeveLine
®
1
0x0000
Temperature (°C × 100)
2
0x0001
Barometric pressure (mbar × 10) – upper 16 bits
3
0x0002
Barometric pressure (mbar × 10) – lower 16 bits
Standard LeveLine
®
Modbus Input Registers
Register
Number
Register
Address (used
in Modbus
commands)
Description: LeveLine – CTD
®
1
0x0000
Temperature (°C × 100)
2
0x0001
Pressure (mbar × 10) – upper 16 bits
3
0x0002
Pressure (mbar × 10) – lower 16 bits
4
0x0003
Electrical conductivity (μS/cm) – upper 16 bits
5
0x0004
Electrical conductivity (μS/cm) – lower 16 bits
6
0x0005
Electrical conductivity (corrected to 20°C) (μS/cm) – upper
16 bits
7
0x0006
Electrical conductivity (corrected to 20°C) (μS/cm) – lower
16 bits
8
0x0007
Electrical conductivity (corrected to 25°C) (μS/cm) – upper
16 bits
9
0x0008
Electrical conductivity (corrected to 25°C) (μS/cm) – lower
16 bits
10
0x0009
Salinity (PSU/ppt × 100)
LeveLine – CTD
®
Modbus Input Registers
All measurement values are two's complement numbers.
6.15.1. Slave Address
Holding register 1 contains the LeveLine's slave address, which is a value in the range 1-
247. It's factory default value is 1. NOTE: If a command writes a new value to holding
register 1, the slave address contained in the response to the command will be the original
slave address, not the new one.
6.15.2. Modbus Mode
Holding register 2 can be set to select between Modbus RTU and ASCII modes of
communication. Its factory default value is 0 (RTU mode). Note that if this register is set to
a new value, the change in communication mode will occur after the response has been
transmitted.
6.15.3. Serial Communication Speed
Holding register 3 can be set to select the communication baud rate. Its factory default
value is 19200, and other valid values are 1200, 2400, 4800, 9600, 38400 and 57600. Note
that if this register is set to a new value, the change in communication speed will occur after
the response has been transmitted.