6.5.9. Start Concurrent Measurement
This command tells the LeveLine
®
to start making a measurement.
Command
Response
<a>C!
<a><time><nr-values><CR><LF>
•
<time> is the time in seconds (three digits) until the measurement will be ready. The
LeveLine
®
always places a value of “001” in this field.
•
<nr-values> is a two-digit number (“01”-“20”) specifying the number of
measurements that will be returned by a subsequent D0 command. The LeveLine
®
always places a value of “02” in this field.
6.5.10. Additional Concurrent Measurements
The LeveLine
®
will respond to Additional Concurrent Measurements commands, but will not
make any measurements.
Command
Response
<a>C<set>! <a><time><nr-values><CR><LF>
•
<set> is a digit (“1”-“9”) indicating which set of additional measurements is being
requested.
•
<time> is the time in seconds (three digits) until the measurement will be ready. The
LeveLine
®
always places a value of “000” in this field.
•
<nr-values> is a two-digit number (“00”-“20”) specifying the number of
measurements that will be returned by subsequent D commands. The LeveLine
®
always places a value of “00” in this field.
6.5.11. Start Concurrent Measurement and Request CRC
This command tells the LeveLine
®
to start making a measurement.
Command
Response
<a>CC!
<a><time><nr-values><CR><LF>
•
<time> is the time in seconds (three digits) until the measurement will be ready. The
LeveLine
®
always places a value of “001” in this field.
•
<nr-values> is a two-digit number (“01”-“20”) specifying the number of
measurements that will be returned by a subsequent D0 command. The LeveLine
®
always places a value of “02” in this field.