LeveLine
®
Instruction Manual
10401-01026 Rev D
Set value in uS/cm
Then hit OK to cal
Last cal: DD/MMM/YY
Cal value:02570
The third line on the screen tells you the last time the EC was calibrated. The fourth line tells you
the last calibration standard used.
7. Using the four arrow keys to navigate and select, enter the EC Cal value of your calibration
solution
at 25ºC.
If using Aquaread
®
RapidCal™, this value should be 02570.
8. Press the 'OK' key.
If the software detects a problem, an error message will be displayed. If an error message
is displayed, take remedial action and try again.
When calibration is complete, a message will be displayed indicating that the calibration
was successful and a Cell Constant will be displayed.
After calibration, rinse the Probe in fresh water.
4.10.4. EC Cell Constant
The EC cell constant is an indication of the health of the EC cell. Nominally, the EC cell
constant should be around 1.25. This can vary with different calibration standard values
and calibration temperatures.
If the EC cell constant is reported above 2.00 or below 0.5, carefully clean the EC cell
contacts and the inside of the nose sleeve. Also check your calibration solution is fresh and
that you have input the quoted value (at 25ºC) during calibration.
4.10.5. Frequency of EC Calibration
Ideally, EC calibration should be carried out regularly based upon the levels of soiling the
Probe experiences.
On a regular basis, thoroughly clean the inside of the nose sleeve and the four metal EC
contacts situated in the EC cell with a soft cloth or toothbrush and non-abrasive detergent.
Never use solvent or alcohol based products to clean the EC cell
. After cleaning,
replace the nose sleeve and re-calibrate EC.
© 2019 Aquaread
®
Ltd.
www.aquaread.com