6.5.12. Additional Concurrent Measurements and Request CRC
The LeveLine
®
will respond to Additional Concurrent Measurements with CRC commands,
but will not make any measurements.
Command
Response
<a>CC<set>! <a><time><nr-values><CR><LF>
•
<set> is a digit (“1”-“9”) indicating which set of additional measurements is being
requested.
•
<time> is the time in seconds (three digits) until the measurement will be ready. The
LeveLine
®
always places a value of “000” in this field.
•
<nr-values> is a two-digit number (“00”-“20”) specifying the number of
measurements that will be returned by subsequent D commands. The LeveLine
®
always places a value of “00” in this field.
6.5.13. Start Verification
The LeveLine
®
will respond to Start Verification commands, but will not perform any self
tests.
Command
Response
<a>V!
<a><time><nr-values><CR><LF>
•
<time> is the time in seconds (three digits) until the self-test result will be ready. The
LeveLine
®
always places a value of “000” in this field.
•
<nr-values> is a single digit (“0”-“9”) specifying the number of numerical values that
will be returned by subsequent D commands. The LeveLine
®
always places a value
of “0” in this field.
6.5.14. Service Request
The service request is a signal sent by the LeveLine
®
to indicate that a measurement has
been completed and is ready to read.
Service Request
<a><CR><LF>
6.5.15. Send Data
This command is used by the data recorder to retrieve the measurement values after a
measurement has been completed.
Command
Response
<a>D<sequence-
nr>!
<a><values>[<crc>]<CR><LF>
•
<sequence-nr> is a single digit (“0”-“9”) indicating the data packet number being
requested.
•
<values> is a string of measurement results, each consisting of a sign character (“+”
or “-”), 1-7 digits and an optional decimal point. The LeveLine
®
returns the
temperature and pressure in the <values> field after a Start Measurement or Start
Concurrent Measurement (with or without CRC request) if <sequence-nr> is “0”; in
all other cases, the <values> field will be empty.
•
the <crc> field will be present if measurement with CRC was requested, and is a 3
byte string containing the encoded CRC of the response packet.